7. The example screen shows the
Fit6/Path2
status is
XShift Error
. The
XShift
value is not within the defined tolerance. When rectified, the status will change to
OK
.
Figure 7-4: Fault Status Details Screen
8. Examine the data shown on the corresponding
Faults
screen to determine if a
component has failed or if it is operating outside of its correct parameters.
9. Contact your Emerson Customer Care Representative
) if you need further assistance with a
component's Fault Status.
7.2.2
Diagnostics menu screen
The
Diagnostics
screen lists those main components of the analyzer where problems can
be diagnosed using the BIST system. The Diagnostics screen also enables you to check on
the status and, where appropriate, the values of any of the listed components.
The
Diagnostics
menu screen can be accessed from the
Main Menu
screen.
This gives the operator access to a number of options when checking the overall health of
the analyzer.
Procedure
1. Where more options are available than can be displayed on the screen use the
or
keys to scroll up or down onto the next page. The screen will continually wrap
around until the option required is highlighted.
2. Pressing the Forward key
will move to the next screen where available as shown
in the Diagnostics/Config example below. Pressing the screen Refresh key at any
time will refresh the current display screen if required by the operator.
CONFIG
gives the status of the analyzer
CONFIGURATION
file.
SENSOR
gives the status of the analyzer
SENSOR(s)
engine.
Troubleshooting and diagnostics
Reference Manual
January 2020
00809-0100-4244
70
Rosemount CT4400
Содержание Rosemount CT4400
Страница 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Страница 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Страница 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Страница 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Страница 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Страница 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Страница 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Страница 105: ......