9. Scroll
or
and select
RUN SPAN
. Press
.
Figure 5-10: For NH3
Note
Allow the concentration readings to stabilize.
10. Use
or
to highlight each digit in turn; then use
or
to increase or
decrease the value until the concentration displayed matches the cylinder you are
using. Once the concentration has been set, Press
to accept.
Figure 5-11: Verify Input
11. The
VERIFY SPAN
screen will be visible while the analyzer makes the verification
measurement.
Allow the concentration readings to stabilize. After the purge time set has elapsed,
the RESULT screen will then be shown.
Verifying gas concentrations
Reference Manual
January 2020
00809-0100-4244
46
Rosemount CT4400