Figure 5-12: Verify Span
12. This
RESULT
screen gives a reading of the concentration for the selected gas that is
present as an impurity in the calibration gas. In the example the gas being
measured is NH
3
, and the analyzer has detected a concentration of 0.510 ppm.
Figure 5-13: Result Screen
If the reading is within tolerance, no further action is required.
If the reading is outside tolerance, follow the zero calibration procedure in
13. Press
to end the span verification process
Reference Manual
Verifying gas concentrations
00809-0100-4244
January 2020
Reference Manual
47