WARNING
HAZARDOUS SUBSTANCES
The analyzer may contain hazardous substances. Always handle the analyzer assemblies
and components with extreme caution.
Gas handling components within the analyzer will contain particulate matter residue from
the sample gases. Over the life of the analyzer, the concentration of the particulate matter
will become enriched within the gas handling components.
When performing repairs or maintenance on the analyzer:
—
Handle used gas handling components with extreme caution.
—
Avoid direct skin contact with used gas handling components.
—
Do not smoke, drink, or eat in the work area.
—
Wear goggles or eye shields.
—
Wear a suitable face mask to protect against inhalation of particulate matter.
—
Do not wet fingers, eyes, or any exposed skin.
—
Pack used gas handling components for disposal in sealed packaging and label them
Contaminated
.
—
Dispose of contaminated items as hazardous material according to the applicable local,
national, or international health and safety regulations and pollution regulations.
7.2
Using the Built-In-Self Test (BIST) fault
diagnostics
The analyzer has a Built-In-Self-Test (BIST) function that can be used to perform failure
diagnosis of some functions other than a complete failure of the equipment. The BIST is
accessed and controlled through the display controller mounted on the front panel of the
analyzer. The BIST runs in the background when the analyzer is operating and continuously
monitors the analyzer for faults.
If the BIST detects a fault the
STATUS
display in the bottom left-hand corner of the
Main
Screen
OK
to a flashing .
Analysis of the fault detected can be performed using the indicated
screen or
.
7.2.1
Fault menu
The BIST function can be used to perform failure diagnosis of some functions in situations
where there is a fault other than a complete failure of the equipment. The BIST is accessed
and controlled through the
FAULTS
screen accessed from the
Main Menu
screen mounted
on the front panel. This gives the operator access to any assess any Faults that the analyzer
has detected in the systems being monitored.
Reference Manual
Troubleshooting and diagnostics
00809-0100-4244
January 2020
Reference Manual
67
Содержание Rosemount CT4400
Страница 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Страница 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Страница 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Страница 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Страница 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Страница 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Страница 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Страница 105: ......