Configuring LOI password (AMS Device Manager)
1.
Right click on the device and select
Configure.
2.
Select
Manual Setup
, and then select the
Security
tab.
3.
Within the Local Operator Interface box, click the
Configure Password
button.
4.
Follow the screen prompts:
a.
Enable the LOI password protection.
b. Enter a 4-digit number as the password.
Configuring LOI password (using Local Operator Interface)
1.
Press any LOI button to activate the menu.
(See
for assistance with using the LOI buttons).
2.
Scroll down (
) and then select
EXTENDED MENU
(
).
3.
Scroll down (
) and then select
PASSWORD
(
).
4.
Enable the LOI password protection.
5.
Enter a 4-digit number as the password.
6.
Exit the menu system by either waiting one minute for the
EXIT MENU?
prompt, or
scrolling down the menus to find and select
BACK TO MENU
and
EXIT TO MENU
.
2.15
Setting the alarm switch
On the electronics board is an alarm switch (see
Follow the steps below to configure the alarm switch.
1.
Set process loop to manual and remove power.
2.
Remove the housing cover.
3.
Use a small screwdriver to slide the alarm switch to desired position.
4.
Replace the housing cover.
5.
The cover must be fully engaged to comply with explosion-proof requirements.
Содержание Rosemount 2140
Страница 1: ...Reference Manual 00809 0100 4140 Rev AA January 2017 Rosemount 2140 Level Detector Vibrating Fork ...
Страница 2: ......
Страница 8: ...VI Reference Manual 00809 0100 4140 Rev AA Contents January 2017 Contents ...
Страница 14: ...4 Reference Manual 00809 0100 4140 Rev AA Introduction January 2017 Introduction ...
Страница 112: ...102 Reference Manual 00809 0100 4140 Rev AA Troubleshooting January 2017 Troubleshooting ...
Страница 143: ......