![background image](http://html.mh-extra.com/html/emerson/ams-trex/ams-trex_user-manual_101320027.webp)
2.6.7
Remove the power module from the Trex unit
CAUTION!
•
Ensure sufficient grounding. Ensure the personnel, working surfaces, and packaging are
sufficiently grounded when handling electrostatically sensitive parts.
•
Avoid touching the pins on the connectors or components. Discharged energy can affect
the modules.
WARNING!
Do not install, remove, or charge the Lithium Ion (Li-Ion) power module in a hazardous area
environment.
Procedure
1.
Turn off the Trex unit.
2.
Place the Trex unit face down on a level, secure surface.
3.
Use a Torx
®
screwdriver to loosen the four screws on the power module. The screws
do not need to be fully removed.
4.
Gently lift the power module straight away from the Trex unit. Do not slide the
power module.
Trex unit with the power module removed
Figure 2-10:
2.6.8
Maintain the power module
To help maintain the performance and life of the power module, understand and follow
the guidelines below:
•
Recharge the power module frequently, preferably after each use or at night. Limit
the number of full discharges, if possible.
•
Avoid frequent use at high temperatures because this can reduce performance.
AMS Trex Device Communicator overview
User Guide
19
Содержание AMS Trex
Страница 1: ...User Guide Rev 3 August 2017 AMS Trex Device Communicator User Guide ...
Страница 8: ...Contents vi User Guide ...
Страница 130: ...Field Communicator application 122 User Guide ...
Страница 186: ...Fieldbus Diagnostics application 178 User Guide ...
Страница 198: ...Technical specifications 190 User Guide ...
Страница 206: ...Glossary 198 User Guide ...