Connection for controlling current
Figure 4-19:
A. Analog input.
2.
On the Loop Diagnostics screen, tap
Trex Unit Current
.
3.
Tap one of the quick buttons to increase the current being supplied, use the up and
down arrow keys, or tap the pencil icon to enter a new value.
4.
Verify the digital control system is reading the same current value.
4.11.4
Simulate a transmitter on an unpowered loop for a loop
check
The Trex unit can provide power and control current to simulate a device for a loop check.
In this setup, the control system is not providing power to the device. This task is used if
the loop will have a 4-wire transmitter.
The Trex unit supplies loop power from the HART + pwr terminals, and Loop Diagnostics
lets you control the current on the loop.
Procedure
1.
Connect the lead set to the
HART + pwr
terminals on the Trex unit and to the
unpowered loop.
Loop Diagnostics application
User Guide
147
Содержание AMS Trex
Страница 1: ...User Guide Rev 3 August 2017 AMS Trex Device Communicator User Guide ...
Страница 8: ...Contents vi User Guide ...
Страница 130: ...Field Communicator application 122 User Guide ...
Страница 186: ...Fieldbus Diagnostics application 178 User Guide ...
Страница 198: ...Technical specifications 190 User Guide ...
Страница 206: ...Glossary 198 User Guide ...