THUM adapter wiring with 250 Ohm resistor
Figure 3-31:
2.
In the Field Communicator application, tap
HART
.
3.
Tap
Yes
when prompted if you want the Trex unit to power the device.
4.
Tap
Positioner
.
5.
When prompted, change the polling address to 63.
6.
Follow any additional prompts to connect to the THUM adapter.
The device menu is displayed.
Related information
HART communication terminals
Wiring diagrams for the Smart Wireless THUM
Adapter and the Field Communicator
application
Disconnect a device
3.13
Connect to an externally-powered HART device
By default, the Trex unit tries to automatically connect to the device at address zero. See
for more information.
CAUTION!
Remove the USB cable from the Trex unit before connecting to a device.
Note
The Trex unit acts as a secondary HART master on a HART loop.
Procedure
1.
Connect the lead set to the HART terminals on the Trex unit and to the
communication terminals on the device. See the HART wiring diagrams topic for the
connection details.
If you want to use the Trex internal resistor, connect the Trex unit is series with the
device.
Field Communicator application
User Guide
81
Содержание AMS Trex
Страница 1: ...User Guide Rev 3 August 2017 AMS Trex Device Communicator User Guide ...
Страница 8: ...Contents vi User Guide ...
Страница 130: ...Field Communicator application 122 User Guide ...
Страница 186: ...Fieldbus Diagnostics application 178 User Guide ...
Страница 198: ...Technical specifications 190 User Guide ...