Only if the atmosphere is not explosive or in de-energised conditions
after a delay of 5 minutes after the electrical supply has been
disconnected, the cover of the housing may be opened!
The cable entries for the electrical connections of the EEx d housing
of the measuring system must never be loosened if the atmosphere is
explosive!
The loosening or unscrewing of the breather is prohibited!
Only authorised personnel is allowed to exchange the breather, if
need be. In this case, the breather as a whole must be exchanged!
Only authorised personnel are allowed to exchange a cable entry
gland, if necessary.
The sensor system may only be switched on before or together with
the
gas-net
Q1 evaluation computer. It will result in an error (alarm) if
you switch on the sensor system after having switched on the