Annex
11
gas-lab
Q1
Page 143
H 2019 T min startup
If during startup of the system, the sensor temperature will not rinse above a minimum
value, this hint is generated along with alarm A606. Meanwhile calculations are not
performed.
H 2020 T instab startup
If during startup of the system, the sensor temperature fluctuates too much this hint is
generated together with alarm A606. Meanwhile calculations are not performed.
H 2021 Sensors off
Information that the gas sensors are shut off. For example, because:
• no pressure
• startup not yet finished
• communication problem between the Q1 sensor unit and controller (together with H423).
H 2022 Path purge
During startup or when the input channel is switched / exchanged, the new channel is
purged first. This is announced by this hint. Meanwhile, there is no measurement.
W753 Calibration timeout
A 1-point calibration has not produced an acceptable (stable) result after a parameterised
time. Hence the calibration is aborted without a result, and the system switches back to
process gas measuring. The warning ends with the next successful calibration. W753 is
always accompanied by A670.
W755 Calibration deviation
A 1-point calibration has produced a result outside the planned confidence interval (1-point
calibration with methane to high). The calibration is aborted without result and the system
switches back to process gas measuring. The warning ends with the next successful
calibration. W755 is always accompanied by A670.