5
Functional description
Page 42
gas-lab
Q1
In case the nitrogen is properly connected to gas path 3 (see beginning of
chapter), please invoke the menu and select the menu item
gas switched on
.
After confirmation the zero point adjustment with nitrogen proceders
automatically.The process takes about 25 minutes. Meanwhile the display
shows the remaining time. It will not count down beforet 2 minutes of flushing
time.
3. Calibration with methane (1. calibration gas)
Now the prepared methane has to be switched on, which the following display
indicates:
Please invoke the menu and select the menu item
gas switched on
.