5
Functional description
Page 44
gas-lab
Q1
After confirming the calibration with H2-11K (2. calibration gas) runs auto-
matically, whereas the remaining time is displayed.
The adjustment starts after approx. 2 minutes for purging and lasts for 15
minutes. When the calibration process with H2-11K (2. calibration gas) is
finished, the following display appears:
Now turn off the H2-11K cylinder and remove the tube on both sides.
(Caution: Unpressurise the line first!).
5. Calibration with L1-8K or binary gas mixture (3. calibration gas)
Now the prepared L1-8K or binary gas mixture has to be connected. (see step 1.
Preparation)Then the prepared 3rd calibration gas is started by opening the
menu and selecting the menu item
gas switched on
.