Installation
7
gas-lab
Q1
Page 113
7.2.2.4
MFE11 input board
The multi-functional MFE11 input board is equipped with 8 digital message
inputs 0/24 V DC. However, it is also possible to use these channels as pulse
inputs with a maximum input frequency of 25 Hz. Connect the inputs at
terminals
D1+
to
D8+
with the common ground
D-
.
Furthermore, the MFE11 input board provides 3 analogue inputs for the
connection of measuring sensors with a 0/4..20 mA output signal. Connect the
inputs via terminals
A1+
/
A2+
/
A3+
with the common ground
A-
.
The analogue inputs of the MFE11 board are electrically isolated from the digital
inputs. Besides, all channels are electrically isolated from the rest of the system.
Connect the cable shielding for the inputs to the
SH
terminal.
7.2.2.5
AE12 input board
The AE12 input board offers 12 analogue inputs for the connection of measuring
sensors with a 0/4..20 mA output signal. Connect the inputs via terminals
I1+
to
I12+
with the common ground
I-
. Connect the cable shielding to the
SH
terminal.