5
Functional description
Page 34
gas-lab
Q1
Manually by performing the
1P-cal. start
command in the
Calibration
menu
what is described below.
By setting a parameterised digital input.
By means of a DSfG command.
The following reactions of the evaluation computer are indicating the status of
the 1-point calibration:
Via measurement outputs and on the controller
’s basic display the last valid
gas quality values are maintained provided any values are available.
The basic display shows
1P-cal.
during calibration.
A
Revision
message will
not be
generated
.
If you want to abort the function prematurely, do so by keeping to the following
instruction sequences:
Invoke the menu from the basic display and select
Calibration
.
Confirm
the
selection
and
invoke
the
menu
again.
Then select
Cal. Cancel
and confirm the selection
The
gas-lab
Q1 terminates the calibration process and returns to normal
operation.