45
Section 4: Troubleshooting and Preventive Maintenance
Error Message
Displayed
Screen
Probable Causes / Possible Solutions
High Resistance
Fault
Moisten Skin
Iontophoresis
Probable Causes:
No Pilogel discs under electrodes.
Dirty electrodes.
Electrodes not properly latched with locking ring or cable
becoming detached from Nanoduct.
Holders not adequately secured to patient.
Damaged electrode cable assembly.
Patient’s skin resistance is too high for proper
iontophoresis current.
Possible Solutions:
Ensure that a Pilogel disc is under each electrode.
Clean electrodes if necessary. Place a drop of deionized
water directly on the clean skin beneath each Pilogel disc.
Ensure both electrodes are latched with locking ring and
that the cable is properly connected to the Nanoduct.
Ensure holders are secured with adequate tightness to
the patient’s limb. Do not overtighten and impede blood
circulation.
Inspect the electrode cable assembly for any damage.
Do not use electrodes if there is any damage to the cable
insulation or if there are exposed bare wires. Replace
electrode cable assembly.
Repeat the sweat test one time.†
If fault continues to occur, contact ELITechGroup for further
instructions.
Fault
Check Instrument
Iontophoresis
Probable Causes:
Operational error has been encountered by the system.
Possible Solutions:
Reset the Nanoduct.*
Perform a Self Test on the Nanoduct.
If the problem persists, discontinue use of the Nanoduct and
contact ELITechGroup for further instructions.
Over Current
Fault
Check Instrument
Iontophoresis
Probable Causes:
Iontophoresis current exceeds allowed threshold.
Nanoduct may be damaged. Discontinue use of the
Nanoduct.
Possible Solutions:
Contact ELITechGroup for further instructions.