Section 4: Troubleshooting and Preventive Maintenance
Error Message
Displayed
Screen
Probable Causes / Possible Solutions
Open Loop
Fault
Check Electrodes
Iontophoresis
Probable Causes:
No Pilogel discs under electrodes.
Dirty electrodes.
Electrodes not properly latched with locking ring or cable
not connected to Nanoduct.
Holders not adequately secured to patient.
Damaged electrode cable assembly.
Possible Solutions:
Ensure that a Pilogel disc is under each electrode.
Clean electrodes if necessary. Place a drop of deionized
water directly on the clean skin beneath each Pilogel disc.
Ensure both electrodes are latched with locking ring and
that the cable is properly connected to the Nanoduct.
Ensure holders are secured with adequate tightness to
the patient’s limb. Do not overtighten and impede blood
circulation.
Inspect the electrode cable assembly for any damage.
Do not use electrodes if there is any damage to the cable
insulation or if there are exposed bare wires.
Repeat the sweat test one time.†
If fault occurs on a regular basis, contact ELITechGroup for further
instructions.