43
Section 4: Troubleshooting and Preventive Maintenance
Error Message Troubleshooting and Diagnosis
Error Message
Displayed
Screen
Probable Causes / Possible Solutions
Low Battery
Main Menu
Probable Causes:
Battery voltage is below the low battery threshold.
Possible Solutions:
Replace all four batteries.
After replacing the batteries, if the same problem persists, contact
ELITechGroup for further instructions.
Fault
Calibration
Not Completed
Check Controls
Probable Causes:
An attempt to calibrate to a value other than 80 mmol/L.
Damaged electrode cable assembly or calibration plate.
Possible Solutions:
Repeat calibration using the 80mmol/L position on the
calibration plate.
Inspect the electrode cable assembly for any damage.
Do not use electrodes if there is any damage to the cable
insulation or if there are exposed bare wires.
If fault continues to occur, contact ELITechGroup for further
instructions.