© Edwards Limited 2009. All rights reserved.
Page 25
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical Data
A533-23-880 Issue C
2.6
Electrical data
Electrical supply
Supply voltage
200-208 V at 50/60 Hz, 3-phase
440 V at 60 Hz, 3-phase
Voltage tolerance
10% (except for 208 V at
50 Hz, which are +6%)
Installation Category
Category 2 (EN61010)
Pollution Category
Category 1 (EN61010)
Full load current ratings
refer to Table 5
Fuse types and ratings
F1
*
*
These fuse rating are included for information only. You cannot change these fuses; a
Edwards service engineer must change these fuses.
30 A, 600 V, Class J
F2
30 A, 600 V, Class J
F3
30 A, 600 V, Class J
F4
20 A, 600 V, Class J
F5
20 A, 600 V, Class J
F6
20 A, 600 V, Class J
F7
1.6 A, 250 V
F8
1.6 A, 250 V
F9
3.15 A, 250 V
Emergency stop switch rating
24 V a.c./d.c., 5 A max
External emergency stop switch
Rating
24 V a.c., 5 A
Load
700 mA (inductive)
Table 5 - Full load current ratings
Supply voltage and frequency
200-208 V
50 Hz
200-208 V
60 Hz
440V
60Hz
FDP
Full load (A)
Motor rating (kW)
22.1
5.5
21.7
5.5
10.1
5.5
HMB
Full load (A)
Motor rating (kW)
14.5
2.9
15.4
3.5
7.2
3.5
Maximum input power to iF system (kW)
11.4
12.1
12.1