D13510840D - SL Sniffer line
Page 44
Maintenance
Service
Whenever equipment is sent back, indicate whether the equipment is contaminated or is free
of substances which could pose a health hazard. If it is contaminated, specify exactly which
substances are involved. You must use the form we have prepared for this purpose.
A copy of this form has been reproduced at the end of these operating instructions: Return
of Edwards Equipment: Declaration (HS2).
Attach the form to the equipment or enclose it with the equipment.
This statement detailing the type of contamination is required to satisfy legal requirements
and for the protection of our employees. We must return to the sender any equipment which
is not accompanied by a contamination statement.
Maintenance work
Sniffer probe clogged
Clogging of the sniffer probe may be due to:
Clogging of capillary filter: refer to
Replacing the felt discs or the capillary filter
Clogging of sinter filter: refer to
Checking/replacing the sinter filter
.
Clogging of sniffer probe capillary
replace sniffer tip.
Damage of sniffer tip
replace sniffer tip.
Clogging/damage of sniffer line
replace ELD500 SL.
WARNING:
Contaminated parts can be detrimental to health and environment.
Before you begin to work, find out whether any parts are contaminated. Adhere to the
relevant regulations and take the necessary precautions when handling contaminated
parts.
CAUTION:
Dirt and damage can impair the function of the vacuum component.
When handling vacuum components, take appropriate measures to ensure cleanliness
and prevent damage.
CAUTION:
Touching the product with bare hands increases the desorption rate.
Always wear clean, lint-free gloves and use clean tools when working in this area.
Содержание ELD500 Series
Страница 9: ...D13510840D Contents Page vii...
Страница 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Страница 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Страница 38: ...D13510840D Page 28 This page has been intentionally left blank...
Страница 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Страница 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Страница 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Страница 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Страница 121: ......