D13510840D - SL Extender interface
Page 37
2. Air helium concentration < 5 x 10
-9
mbar ls
-1
Possible causes:
Incorrect calibration.
Separator clogged.
Clearing the fault:
Recalibrate using a calibrated leak.
Check display with a calibrated leak, if no reaction exchange separator according to
.
3. Air helium concentration > 10
-4
mbar ls
-1
Possible causes:
Incorrect calibration.
Separator faulty or inserted with a leak.
Clearing the fault:
Recalibrate using a calibrated leak.
Exchange separator according to
4. Reaction to a calibrated leak is much too strong or much too weak.
Possible cause:
Incorrect calibration.
Clearing the fault:
Recalibrate; if unsuccessful check calibrated leak, also check causes 2 and 3.
5. No reaction to a calibrated leak.
Possible causes:
Separator clogged.
Filter tip clogged.
Sample line clogged.
Clearing the fault:
Check cause 2.
Exchanging the dust filter in the probe tip
.
Blow out sample line.
Содержание ELD500 Series
Страница 9: ...D13510840D Contents Page vii...
Страница 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Страница 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Страница 38: ...D13510840D Page 28 This page has been intentionally left blank...
Страница 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Страница 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Страница 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Страница 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Страница 121: ......