D13510840D - Interface description
Page 104
Enumerations
LD communication protocol status word
Status word bit no.
ELD500
Bit 0
Device state Bit 0
Bit 1
Device state Bit 1
Bit 2
Device state Bit 2
Bit 3
Sniffer button
Bit 4
ZERO
Bit 5
Still warning
Bit 6
Range 0
Bit 7
Range 1
Bit 8
Range 2
Bit 9
Trigger 1 exceeded
Bit 10
Trigger 2 exceeded
Bit 11
Trigger 3 exceeded
Bit 12
Bit 13
Device Warning
Bit 14
Device Error
Bit 15
Syntax / Command error
Value
Bit 2..0
Device state
0
000
INIT
1
001
RUNUP
2
010
STANDBY
3
011
VENT
4
100
EVACUATION
5
101
MEASURE
6
110
CALIBRATION
7
111
ERROR