D13510840D - Interface description
Page 100
520
0208
Calibration
factors
vacuum
FLOAT[3]
R/W
1.0E-6, 1.0,
1.0E6
Calibration factors
for vacuum mode
Index 0: Mass 2
Index 1: Mass 3
Index 2: Mass 4
521
0209
Calibration
factors sniff FLOAT[3]
R/W
1.0E-6, 1.0,
1.0E6
Calibration factors
for sniff mode
Index 0: Mass 2
Index 1: Mass 3
Index 2: Mass 4
522
020A
Machine
factor
FLOAT
R/W
1.0E-6, 1.0,
1.0E6
Machine factor
523
020B
Factor sniff
FLOAT
R/W
1.0E-6, 1.0,
1.0E6
Factor sniff
530
0212
Cathode
selection
UINT8
R/W
0, 0, 1
Cathode selection
0 = cathode 1
1 = cathode 2
540
021C
Pressure
EVAC GROSS
[mbar]
FLOAT
R/W
3, 15, 15
Pressure limit
EVAC --> GROSS
[mbar]
541
021D
Pressure
GROSS FINE
[mbar]
FLOAT
R/W
0.02, 0.1,
0.1
Pressure limit
GROSS --> FINE
[mbar]
548
0224
Background
suppression UINT8
R/W
0, 1, 2
Background
suppression
0 = disabled
1 = internal
2 = inlet area
550
0226
Vacuum
ranges
UINT8
R/W
0, 0, 5
Vacuum ranges
see table /
enumerations
552
0228
Vent delay
UINT8
R/W
0, 3, 4
Vent delay
0 = immediately
1 = 1 second
2 = 1.5 seconds
3 = 2 seconds
4 = no vent
Command
dec
Command
hex
Name
Data type Read/
Write
Min-, Def-,
Max-value Meaning
Содержание ELD500 Series
Страница 9: ...D13510840D Contents Page vii...
Страница 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Страница 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Страница 38: ...D13510840D Page 28 This page has been intentionally left blank...
Страница 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Страница 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Страница 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Страница 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Страница 121: ......