D13510840D - SL Extender interface
Page 38
6. Response time until a calibrated leak indication occurs is over 2 seconds with a 5 minute
sample line.
Possible causes:
Filter tip clogged.
Sample line clogged.
Clearing the fault:
Exchanging the dust filter in the probe tip
.
Blow out sample line.
Maintenance
When operating the ELD500 SL extender interface for prolonged periods of time under dusty
or damp conditions the tip of the sample probe or the separator may be become
contaminated or clogged. The affected parts will then have to be exchanged.
Any further maintenance is not required especially as the ELD500 SL extender interface
sample pump is maintenance-free.
Exchanging the separator
In case the separator has become contaminated or clogged it can be exchanged as follows:
Detach the ELD500 SL extender interface from the helium-leak detector.
Remove the connection flange (
item 5) on the ELD500 SL extender interface
by loosening the screws (
Exchange the separator (
item 3) and if required also the O-rings
item 2) respectively (
item 4) (all parts must be kept clean, do not
grease).
After having inserted the separator (
item 3) and the O-rings (
item 2) respectively (
item 4) again attach the connection flange (
item 5) by tightening the four screws (
item 6) (tighten screws clockwise).
Attach the ELD500 SL extender interface to the leak detector (see
).
Check operation as described in
.
Содержание ELD500 Series
Страница 9: ...D13510840D Contents Page vii...
Страница 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Страница 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Страница 38: ...D13510840D Page 28 This page has been intentionally left blank...
Страница 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Страница 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Страница 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Страница 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Страница 121: ......