D13510840D - SL Extender interface
Page 35
Figure 33 Connection of the SL extender interface to the ELD500 leak detector
Initial operation
1. Connect to the mains according to
Mains connection and exchanging the ELD500 SL
2. Make the vacuum connection, according to
.
3. Switch on the ELD500 SL extender interface via mains switch (
4. Switch on the helium leak detector.
5. After switching on the helium leak detector wait for the unit to run-up until the Stand-by
mode is reached.
By operating the pushbutton "PUMP/START" pump the ELD500 SL extender interface down
until it is ready for operation. This is indicated by:
transition to leak rate indication and
indication of an air helium background.
Function test
1. Suppress the helium background by pressing the zero button at the leak detector.
2. Press "TEST" pushbutton (
item 5) on the ELD500 SL extender interface; the
indication will then change from zero to several 10
-9
mbar ls
-1
.
1. Housing of the ELD500 SL
extender interface
4. Connection flange of the ELD500
SL extender interface
2. Housing of the ELD500
5. O-Ring
3. Conical ring (metal)
6. Connection flange of the ELD500
Содержание ELD500 Series
Страница 9: ...D13510840D Contents Page vii...
Страница 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Страница 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Страница 38: ...D13510840D Page 28 This page has been intentionally left blank...
Страница 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Страница 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Страница 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Страница 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Страница 121: ......