D13510840D - Interface description
Page 103
1630
065E
Background
calculation
[A]
FLOAT
R
Results inlet area
background
calculation
Index 0 = Offset
internal
Index 1 = Offset
external old
Index 2 = Offset
external new
1854
073E
Contaminati
on limit
[mbar*l/s]
R/W
FLOAT
1.0E-6,
1.0E-3,
1.0E3
Contamination
limit
1855
073F
Contaminati
on protect
R/W
UINT8 0, 0, 1
Contamination
protection
0 = disabled
1 = enabled
2500
09C4
Last
parameter
set
CHAR[*]
R
Last parameter set
see table /
enumerations
2501
09C5
Parameter
set
UINT8
R/W
Parameter set
0 = load default
1...3 = load 1...3
4...6 = save 1...3
2591
0A1F
Control
location
UINT8
R/W
0, 0, 5
Control location
0 = local
1 = RS232
2 = PLC
3 = local / RS232
4 = local / PLC
5 = local / PLC /
RS232
2660
0A64
Maintenance
activ
UINT8
R/W
0, 0, 1
Service message
oil filter
0 = disabled
1 = enabled
Command
dec
Command
hex
Name
Data type Read/
Write
Min-, Def-,
Max-value Meaning
Содержание ELD500 Series
Страница 9: ...D13510840D Contents Page vii...
Страница 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Страница 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Страница 38: ...D13510840D Page 28 This page has been intentionally left blank...
Страница 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Страница 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Страница 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Страница 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Страница 121: ......