D13510840D - SL Sniffer line
Page 45
Figure 38 Sniffer tip
Replacing the felt discs or the capillary filter
Switch the leak detector off or disconnect the sniffer line.
Remove dirty felt discs by means of tweezers and replace them.
or
unscrew capillary filter and replace it.
(If using the metallic version, do not forget the seal.)
Checking/replacing the sinter filter
Switch the leak detector off or disconnect the sniffer line.
Remove the two Phillips screws.
Remove the sinter filter with the seal.
Visually check the filter for contamination:
1. Capillary filter (plastic version; standard)
4. Phillips screws
2. Felt discs
5. Sinter filter
3. Capillary filter
(Metallic version with seal; option)
6. Seal