Ser in u s ® 30 U ser M an u al 3 .3
Page 84
3.4.28
Digital Output Menu
Main Menu
→
Communications Menu
→
Digital Output Menu
This allows the instrument to trigger external alarms in response to certain events. There are eight
different pins available, which will be set high during an associated event:
Figure 70
–
Digital Output Menu Screen
DO N (Pin X)
Associates a state with a digital output. There are eight digital
outputs (the pin numbers are for the 25 pin connector). Each one
can have one of the associated states listed in Table 6. The pin
will be driven to the active state while the instrument state is
true.
Active
Each pin can be set to be active
High
or
Low
. Active High means
that the pin will be pulled to 5 V when the associated event
occurs. Active Low means the pin will be pulled to 0 V when the
associated event occurs.
Table 6
–
Digital Output States
Digital Output State
Description
Disabled
No state (this state is never
Active
).
Pwr Supply Fail
Power supply fault.
A2D Fail
Analog to digital fault.
Corr. Wheel Fail
Correlation wheel fault.
Flow Fail
Sample flow fault.
Cell Heat Fail
Cell heater fault.
Conv. Heat Fail
CO Scrubber heater fault.
Flow Heat Fail
Flow block heater fault.
Mirror Heat Fail
Mirror heater fault.
Lamp Heat Fail
Lamp heater fault.
Chassis Tmp Fail
Chassis temperature fault.
USB Disconnected
The USB memory stick is disconnected.
Background
Performing a background.
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...