Optional Extras
Page 177
8.4.5
Flow Calibration
The flow calibration menu is only available when the internal pump option is installed (refer to
Section 3.4.12).
The internal pump requires a separate flow calibration procedure. The flow calibration detailed in
Section 5.8 must be performed after any pump service or pressure calibration.
8.4.6
Internal Pump Leak Check
The internal pump requires a separate modified procedure to leak check the system (refer to
Section 6.4.4.1).
8.4.7
Pump Control
This section describes how to turn OFF and ON the internal pump for servicing and maintenance
purposes.
Procedure
1.
Open - Main Menu
→
Service Menu
→
Diagnostics Menu
→
Internal Pump Menu.
2.
Select -
Pump Control
→
Manual
.
3.
Toggle OFF -
Internal Pump
→
OFF.
8.5
Metric Fittings Kit (PN: E020122)
The metric fittings kit allows the user to connect 6 mm tubing to the rear ports of the analyser. This
can be very handy if it is hard to source
1/4”
tubing from a local supplier.
8.6
High Pressure Zero/Span Valves
High pressure span calibration valve (factory installed (PN: E020108))
High pressure zero calibration valve (factory installed (PN: E020109))
For more information on how to setup and use the high pressure valves refer to Section 5.9.
Refer to Section 9.11 for the exploded assembly drawing.
Note:
Before using a high pressure span or zero as a source for calibrating the instrument, check
local regulatory requirements.
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...