Internal Pump Additional Components
The Serinus 30 internal pump option includes the following additional components:
Table 11
–
Internal Pump Additional Components
Component
Description
Part number
Internal Pump
Draw sample through the instrument
H010027
Flow Block
Includes sintered filter and differential pressure
sensor to measure flow.
H010120
Heater and Thermistor
Mounted in flow block to measure and control
temperature for accurate flow measurement.
Installed in flow
block
Gasket
Used to thermally isolate the flow block assembly
from the Serinus chassis.
H010119
Pressure Cable
Used to connect differential pressure PCA of the flow
block to main controller PCA.
C020062-02
Heater Thermistor Cable
Used to connect heater thermistor assembly of the
flow block to the main controller PCA.
C020083
Stop Valve
Valve used to isolate pump from the pneumatic
circuit of the instrument.
H030170
Micro Wire Saddle
Used to cable management
H030121
8.4.3
Internal Pump Removed Components
The Serinus 30 has a number of components missing from the standard instrument due to the presence
of the internal pump and flow block controlling flow within the instrument. The parts that have been
removed when this option is installed are listed in the table below.
Table 12
–
Internal Pump Removed Components
Component
Part number
Sintered filter
F010004
O-ring
O010012
Spring
H010040
O-ring
O010013
Orifice
H010043-13
8.4.4
Pressure Calibration Internal Pump Option
The internal pump requires a separate modified procedure to allow the internal pump to generate the
necessary vacuum (refer to Section 5.2.3).
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...