Operation
Page 73
Preamp
: Injects an artificial test signal into the
Preamplifier mounted on the
Measurement Cell
to
verify that the Preamplifier, cabling and electronic
circuitry on the main controller PCA is operating
correctly. When in this
Diagnostic Mode
, adjust the
Test Meas Pot
from 0 - 255. This will produce a
change in the concentration voltage as well as the
indicated gas concentration. Adjust the
Test Ref Pot
from 0 - 255 to produce a change in the reference
voltage.
3.4.16
Internal Pump Menu (Option)
Main Menu
→
Service Menu
→
Diagnostics Menu
→
Internal Pump Menu
This menu only appears when the internal pump option is installed. Setting the
Pump Control
to
MANUAL
will automatically disable the valve sequencing.
Figure 58
–
Internal Pump Menu Screen
Flow Set Point
The desired sample flow target.
Cell
The current pressure in the measurement cell.
Sample Flow
Current gas flow. This is only accurate when reading close to the
flow calibration point.
Internal Pump
This field allows the internal pump to be turned On or Off. This
field is only editable when the
Pump Control
field is set to
MANUAL
.
Pump Control
Set to
MANUAL
to disable the automatic pump control.
AUTO
allows the flow PID to modify the pump coarse and fine settings.
START
will transition to
AUTO
after one second.
Coarse
Internal pump speed control (
Coarse
). This field is only editable
when the
Pump Control
field is set to
MANUAL
.
Fine
Internal Pump speed control (
Fine
). This field is only editable when
the
Pump Control
field is set to
MANUAL
.
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...