Ser in u s ® 30 U ser M an u al 3 .3
Page 216
Appendix B.
EC9800 Protocol
The Serinus implements a subset of the 9800 instrument protocol. Only the basic commands of reading
the concentration value and setting the instrument calibration state (measure, span or zero) are
supported.
B.1
Command Format
All commands are sent as ASCII strings. Fields are delimited by commas and the command ends with
the normal return key (i.e. the TERMINATOR is either a <CR> or a <LF>). The DEVICE I.D. is the Serial ID
assigned in the
Main Menu
→
Communications Menu
→
Serial Communication Menu
. If the
instrument is not being used in a multi-drop connection, the DEVICE I.D> can be replaced with the
string “???”.
B.2
Commands
DCONC
Function:
Sends the current instantaneous concentration data to the serial port.
Format:
DCONC, {<DEVICE I.D.>} {TERMINATOR}
Device response:
{GAS} <SPACE> {STATUS WORD} <CR><LF>
The GAS value is the concentration value in the current instrument units, expressed as a floating point
number (i.e. 12.345). The STATUS WORD indicates the instrument status in hex (i.e. A01F) using the
following format:
Bit 15 = SYSFAIL (MSB)
Bit 14 = FLOWFAIL
Bit 13 = LAMPFAIL
Bit 12 = CHOPFAIL
Bit 11 = CVFAIL
Bit 10 = COOLERFAIL
Bit 9
= HEATERFAIL
Bit 8
= REFFAIL
Bit 7
= PS-FAIL
Bit 6
= HV-FAIL
Bit 5
= OUT OF SERVICE
Bit 4
= Instrument is in zero mode
Bit 3
= Instrument is in span mode
Bit 2
= Unused
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...