Installation/Decommission
Page 41
Note:
The power ON/OFF switch is accessible from the rear of the instrument only. Install
the instrument so that the ON/OFF power switch is accessible.
2.3
Instrument Set-Up
After installing the instrument, the following procedures should be followed to ready the instrument
for monitoring:
Figure 27
–
Instrument Rear Panel
2.3.1
Power Connections
CAUTION
Hazardous voltages exist within the instrument. Do not remove or modify any of
the internal components or electrical connections whilst the mains power is ON.
CAUTION
Always unplug the equipment prior to removing or replacing any components.
CAUTION
Do not replace the detachable mains supply cord with an inadequately rated
cord. Any mains supply cord that is used with the instrument must comply with
the safety requirements (250 V/10 A minimum requirement).
CAUTION
Ensure that the mains supply cord is maintained in a safe working condition.
CAUTION
When connecting the mains power to the instrument, the following must be
adhered to otherwise the safety and the reliability of the instrument may be
compromised.
A three pin mains power cord with a protective earth conductor
MUST
be used.
The mains power outlet (wall socket) must be in the range of 100 - 240 VAC, 50 - 60 Hz.
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...