Appendix
Page 219
Note: The I.D. of 97 is sent as the sequence 097. All I.D. strings must have three digits and must
always be padded with ASCII zero characters.
This is an example of a valid command to put the unit in the manual span mode if the instrument has
an ID of 843:
<STX>ST843 K<ETX>52
The block check operation is best shown with the following table:
Table 35
–
Block Check Operation
Character
Hex Value
Binary
Block Check
<STX>
02
0000 0010
0000 0010
S
53
0101 0011
0101 0001
T
54
0101 0100
0000 0101
8
38
0011 1000
0011 1101
4
34
0011 0100
0000 1001
3
33
0011 0011
0011 1010
<SPACE>
20
0010 0000
0001 1010
K
4B
0100 1011
0101 0001
<ETX>
03
0000 0011
0101 0010
The binary block check value is 0101 0010 which is the hex value 52 as shown at the end of the
command string.
C.2
Commands
DA
Return the current instantaneous concentration.
Command Format
<STX>{DA}{<kkk>}<ETX>< bcc1><bcc2>
Where:
kkk
Device’s ID. This field is optional, but if provided it must be padded with
zeros to be 3 characters long. The value must match one of the
following: the instrument’s Bayern
-Hessen ID, 000, or ??? (three
question marks).
bcc1
First byte of the block check calculation.
bcc2
Second byte of the block check calculation.
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...