Serial Delay
Some older communication systems require a delay before
the instrument responds to a serial command. The number
of milliseconds of delay required (0 - 1000). The default is 0,
meaning the instrument responds as quickly as possible to
any serial request.
Baudrate
Sets the baud rate for this serial port (
1200
,
2400
,
4800
,
9600
,
14400
,
19200
,
38400
or
115200
).
Protocol
Sets the protocol used for this serial port (
Advanced
,
ModBus
,
EC9800
or
Bayern-Hessen
). This must be set to
Advanced
for Ecotech supplied software.
Endian
[Modbus Protocol]
Select
Little
or
Big
endian mode for
ModBus
protocol.
3.4.25
Analog Input Menu
Main Menu
→
Communications Menu
→
Analog Input Menu
The Serinus supports three analog inputs from the analog & digital I/O port. Each input is a 0 - 5 volt
CAT 1 input that can be scaled and logged to the USB memory stick or accessed remotely as parameters
199 - 201.
CAUTION
Exceeding these voltages can permanently damage the instrument and void the
warranty.
Figure 67
–
Analog Input Menu Screen
Input 1 is CO2
When enabled several new menu items appear in other menus
and the home screen now displays the CO and CO
2
gas
concentrations.
Input 1/2/3
The sections below are repeated for each analog input.
Multiplier
The input voltage will be multiplied by this number. For
example, if a sensor has a 0 - 5 V output for a temperature of
-40 °C to 60 °C, the multiplier would be (60 - (-40))/5 = 20.
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...