Ser in u s ® 30 U ser M an u al 3 .3
Page 208
75
Cell Temperature
Cell Temperature
76
Conv. Temperature
Converter Temperature
77
Chassis Temperature
Chassis Temperature
78
Manifold Temp.
Manifold Temperature
79
Cooler Temperature
Cooler Temperature
80
Mirror Temperature
Mirror Temperature
81
Lamp Temperature
Lamp Temperature
82
O3 Gen. Lamp Temp.
O3 Lamp Temperature
83
Instrument Status
Each bit in this 4-byte word represents a different condition
(not all conditions apply to every instrument model):
BIT
Condition if set
0
Currently in warmup process
1
Volumetric units (ppm); otherwise gravimetric units
(mg/m3)
2
Performing a background
3
Currently in Span mode
4
Currently in Zero mode
5
Instrument Out of Service (or in Diagnostic mode,
PTF compensation or control loop disabled, or
Comms debugging enabled)
6
High Voltage failure
7
System power failure (not actually possible to
report)
8
Reference voltage failure
9
Cell temperature failure
10
Cooler failure
11
Converter failure
12
Correlation wheel failure
13
Lamp source failure
14
Flow fault
15
Any system error (the red instrument panel light is
ON)
84
Reference Voltage
Units in Volts
85
Calibration State
This variable has two different sets of values:
Set Calibration State
0 = MEASURE
1 = CYCLE
2 = ZERO
Get IEEE Value
0 = MEASURE
1 = ZERO
2 = SPAN
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...