Communications
Page 97
6.
Select -
Protocol
→
Advanced
- Accept.
Figure 81
–
Example of Network Menu Setup
7.
Once completed, use the power switch on the rear of the instrument to turn the power OFF. Leave
the instrument OFF for 10 seconds before turning the power back ON.
Note:
Manually perform a hardware power cycle every time the IP address is changed for it to take
effect.
8.
Open -
Main Menu
→
Communications Menu
→
Network Menu
.
9.
The
Start-up Mode
automatically changes to
Read IP
and the current network port settings will be
displayed on the screen.
10.
When viewing is complete select -
Start-up Mode
→
Normal
- Accept.
4.3.3
Port Forwarding on Remote Modem/Router Setup
When using the network port to connect to the router / modem with NAT enabled, the user will need
to add IP mapping to ensure that data is forwarded through to the desired port. This is known as port
forwarding. To set-up the port for the instrument, the user needs to go into the modem/router
configuration. Normally, the user will see the port forwarding setup under Port Forwarding, NAT, or
Port Mapping menu. Below is an example port forwarding setup .
The default port for the Serinus range of instruments is
32783
. The destination address is the
instrument IP address setup in the
Network Menu
.
Figure 82
–
Port Forwarding Example
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...