Operation
Page 69
Ambient
The current ambient pressure.
Cell
The current pressure in the measurement cell.
The current measurement cell pressure displayed as a raw voltage.
Vacuum Cal Mode
Defaults to Off. When turned On, the valves will be set to the same
state as during a
Vacuum Set Pt.
adjustment, but there is no
adjustment. Used for checking the accuracy of the vacuum
pressure calibration. There are no dialog boxes or prompts, so the
user needs to follow similar steps and precautions as during
Vacuum Set Pt.
Ambient Cal Mode
Defaults to OFF. Similar to
Vacuum Cal Mode
, except the valves are
set to check the ambient calibration.
3.4.12
Flow Calibration Menu (Option)
Main Menu
→
Calibration Menu
→
Flow Calibration Menu
This menu only appears when the internal pump option is installed. Refer to Section 5.8 for the
calibration procedure. Setting the
Pump Control
to
MANUAL
will automatically disable the valve
sequencing ready for flow calibration.
Figure 54
–
Flow Calibration Menu Screen
Flow Set Point
The desired sample flow target.
Cell
The current pressure in the measurement cell.
Cal. Point
The flow at which the last flow calibration was performed. If
Pump
Control
is set to
MANUAL
and
Internal Pump
is
ON
, this field can be
edited to calibrate the current flow against a reference.
Calibration must be done at or near the flow set point for best results
(refer to Section 5.8).
Cal. Zero
If
Pump Control
is set to
MANUAL
and
Internal Pump
is
OFF
,
activating this command will calibrate the flow sensor zero point (refer
to Section 5.8).
Sample Flow
Current sample flow through the instrument. This is only accurate
when reading close to the flow calibration point.
Internal Pump
This field allows the internal pump to be turned ON or OFF. This field is
only editable when the
Pump Control
field is set to
MANUAL
.
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...