26
zero calibration, or it can reset both alarm and peak and perform remote zero calibration. To verify
this parameter or to change it, press the
FUNC
key until
nonE
and
GROUP
show on the display.
Press the
▲
key until 4 shows in the upper display. Press the
FUNC
key until the lower display
shows LI.TYP. Press the
▼
or
▲
key until the upper display shows the correct selection: OFF, AL -
Alarms Reset, P - Peak Reset, AL-P - Alarm & Peak Reset, CAL.0 - Zero Calibration or ALL - All
Functions. Press the
FUNC
key to set the value and move to the next parameter, or press the RESET
key to go back to the active display.
8.3.2 S
ETTING
THE
L
OGIC
I
NPUT
S
TATUS
(
IF
SUPPLIED
)
The Logic Input Status can be off, can be set to function as an alarm reset, a peak reset, or it can
reset both. To verify this parameter or to change it, press the
FUNC
key until
nonE
and
GROUP
show on the display. Press the
▲
key until
4
shows in the upper display. Press the
FUNC
key until
the lower display shows
LI.STS
. Press the
▼
or
▲
key until the upper display shows the correct
selection:
CLOSE
, or
OPEn
.
Press the
FUNC
key to set the value and move to the next parameter, or press the
RESET
key to go
back to the active display.
8.3.3 S
ETTING
P
EAK
D
ETECTION
The Peak Detection can be either set to OFF, the default value of HIGH, or to LOW. To verify this
parameter or to change it, press the
FUNC
key until
nonE
and
GROUP
show on the display. Press
the
▲
key until
4
shows in the upper display. Press the
FUNC
key until the lower display shows
PEAK
. Press the
▼
or
▲
key until the upper display shows the correct value (
OFF
,
HI
, or
LO
). Press
the
FUNC
key to set the value and move to the next parameter, or press the
RESET
key to go back
to the active display.
8.3.4 S
ETTING
THE
L
INE
F
REQUENCY
The Line Frequency default value is 60 Hz. To verify this parameter or to change to 50 Hz, press the
FUNC
key until
nonE
and
GROUP
show on the display. Press the
▲
key until
4
shows in the upper
display. Press the
FUNC
key until the lower display shows
LINE.F
. Press the
▼
or
▲
key until the
upper display shows the correct frequency. Press the
FUNC
key to set the value. Press the
FUNC
key to set the value and move to the next parameter, or press the
RESET
key to go back to the
active display.
8.3.5 S
ETTING
THE
D
ISPLAY
F
ILTER
Filtering is an electrical method of averaging the displayed values over a period of time to arrive at a
more legible display. Filtering helps to eliminate short duration transients and spikes that may cause
false or spurious readings.
To change or view the Main Analog Output Filter, press the
FUNC
key until
nonE
and
GROUP
show on the display. Press the
▲
key until
2
shows in the upper display. Press the
FUNC
key until
Содержание UPR700
Страница 14: ...14 Fig 2 Model UPR700 Outline Drawing...
Страница 16: ...16 Fig 3 UPR700 Wiring 4 20mATransmitter Internal 24VDC Power Supply Primary Input...
Страница 33: ...UPR700 Microprocessor Based Pressure Process Indicator 33 Fig 10 Secondary Input Jumper Location...
Страница 67: ...NOTES...
Страница 68: ...NOTES...
Страница 70: ...DYNISCO LLC 38 FORGE PARKWAY FRANKLIN MA 02038 ATTN MARKETING DEPT Place Stamp Here...
Страница 71: ......