UPR700 Microprocessor-Based Pressure/Process Indicator
19
NOTE 3: Relay Outputs
The contact rating of all outputs is equal to 2A/240 Vac on resistive load.
•
To avoid electrical shock, connect power line at the end of the wiring procedure.
•
For power connections use No 16 AWG or larger wires rated for at least 75
°
C.
•
Use copper conductors only.
NOTE 4: Power Line
Before connecting the instrument to the power line, make sure that the line voltage corresponds to
the description on the identification label.
To avoid electrical shock, connect power line at the end of the wiring procedure. For supply
connections use No. 16 AWG or larger wires rated for at least 75
°
C.
•
Use copper conductors only.
•
Don’t run input wires together with power cables.
•
For 24 V DC the polarity need not be observed.
The power supply input is fuse protected by a sub miniature fuse rated T, 1A, 250 V. When the fuse
is damaged, it is advisable to verify the power supply circuit It is necessary to send back the
instrument to Dynisco for service.
The safety requirements for Permanently Connected Equipment say:
•
A switch or circuit-breaker shall be included In the building installation;
•
It shall be in close proximity to the equipment and within easy reach of the operator
•
It shall be marked as the disconnecting device for the equipment
NOTE 5:
A single switch or circuit breaker can drive more than one Instrument:
•
When a neutral line is present, please connect it to terminal 54.
•
Protective conductor terminals shall be connected to earth.
Содержание UPR700
Страница 14: ...14 Fig 2 Model UPR700 Outline Drawing...
Страница 16: ...16 Fig 3 UPR700 Wiring 4 20mATransmitter Internal 24VDC Power Supply Primary Input...
Страница 33: ...UPR700 Microprocessor Based Pressure Process Indicator 33 Fig 10 Secondary Input Jumper Location...
Страница 67: ...NOTES...
Страница 68: ...NOTES...
Страница 70: ...DYNISCO LLC 38 FORGE PARKWAY FRANKLIN MA 02038 ATTN MARKETING DEPT Place Stamp Here...
Страница 71: ......