This model is a single bore rheometer and has Load Cell measurement capability only. You
will need to install the Dynisco Load Cell to the rheometer crossbar and connect the coiled
interface cable from the back of the rheometer base to the Load Cell. (See Load Cell
Assembly Section for more details).
This model is a single bore rheometer that has simultaneous Load Cell and/or PT
measurement capability. You will need to install the Dynisco Load Cell to the cross bar, and
install the Dynisco PT into the rheometer barrel. Connect the coiled interface cable from
the rheometer drive system at the top of the rheometer to the Load Cell, and another coiled
interface cable from the back of the rheometer base to the PT. (See Load Cell Assembly and
Pressure Transducer Installation sections bellow for more details).
This model rheometer has a dual bore and Twin Pressure Transducers for twin die
measurement capability. Bagley and entrance pressure loss measurements can be
performed in one test using this instrument. Install both of the Dynisco PT’s into the
rheometer barrel and connect the coiled interface cable from the rheometer base to the
PT’s. (See Pressure Transducer Installation Section for more details). Be sure Sensor #1 is
attached to barrel 1, and Sensor #2 is attached to barrel 2.
Models 7000 & 7001 have Load Cell measurement capability. The Load Cell is a round, blue
unit with a 6-pin BENDIX connector. You will need to mount the Load Cell to the crosshead
and connect it to the rheometer’s coiled interface cable. The actual deflection (which
measures the force) occurs on the side of the cell, which has the exposed bolt heads. The
opposite end is considered the mechanical ground. Screw the plunger load cap (hollow-cup
shaped with screw threads on one end) into the side of the load cell that has the exposed
bolt heads (active end). The rheometer plunger will fit into the load cap when a test is
underway. Connect the load cell to the cross-head of the rheometer as follows:
Screw the open end onto the bolt in the center of the crosshead located between the two-
rheometer posts and directly above the barrel hole. A two handed snug tightening is all that
is required. Connect the coiled interface cable to the Load Cell.
7.
Model 7000
8.
Model 7001
9.
Model 7002
10.
Load Cell Assembly (models 7000 & 7001)