Immediately prior to pressing RUN
Plunger in the rheometer?
Guide Bushing in place?
Swing arm fully closed?
Area between crosshead and swing arm clear?
Die Nut tight? Pressure transducer tight?
Prior to loading Sample
Is material properly prepared (dried, mixed, check for contaminates)
Rheometer on 20 minutes?
Correct Die in rheometer?
Sufficient anti-seize coating on die nut
(especially at high temperatures <300
°
C).
Correct program being run?
Hand tools in position (packing funnel and tool, cleaning drill bit etc.)
Are die diameters within spec. (passed G0-No Go gage, ASTM, ISO, DIN?)
Force calibration and R-Cal OK?
Temperature calibration OK?
Die length OK?
Piston Tip Diameter within spec.?
Threads OK on die holder nut? (Too little anti-seize causes excessive wear)
Barrel Diameter OK?
6.
Generic Pre-Run Check List