LCR7000 & RHEOSIGHTS P/N: 974194 |
Rev: A
|
ECO: 52468
www.dynisco.com
From lab to production,
providing a window into the process
-73-
FOR 7002 SKIP TO 3.2.
Load cell (LC) configuration and calibration.
B.
C
ALIBRATING THE
R
HEOMETER
L
OAD
C
ELL WITH THE
C
ALIBRATION
U
NIT
Connect load cell sensor.
Carefully align 6 pin Bendix female connector of the 7052-37 to the 6 pin male
connector of the load cell.
Enter device into library.
Press
SYS
on keypad.
Press
DEVICE
smart key.
Press
LC
smart key.
Press
until desired LC# (load cell sensor number) is viewed on display.
Press
NEW
smart key.
Press
until desired load cell sensor is viewed on delay.
Press
ENTER
smart key.
Press
. LC UNIT is displayed and should read: N.
Press
. LC CAPACITY is displayed and should read: LC capacity.
Press
. LC SCALING is displayed and should read: 2.000 mV/V.
Press
. GAIN is displayed and should read: 128.
Press
. LC R-CAL is displayed and should read: 72.920%.
Press
. LC SEQUENCE is displayed and should be blank.
Press
ESC
once on keypad.
Press
SAVE
smart key.