Press
SYS
key and
CAL
smart key to enter calibration mode.
Select RTD Sensor in the Calibrate Select mode.
Press
SELECT
smart key.
Select
SINGLE
. Press
STABLE
smart key. Press
ENTER
then press
SAVE
. This sets offset to
0.
Allow unit to re-stabilize for 10 minutes.
Press
SYS
key and
CAL
smart key. Select RTD Sensor. Select
DOUBLE
.
Press
STABLE
smart key.
Enter reading displayed on the BTP 1000A display.
Use the number pad on the right side of the keypad to enter the temperature. Enter the
numbers in order as they appear on the BTP 1000A display, also pressing the decimal
where necessary. NOTE: numbers are entered by just pressing the appropriate number
key.
Press
ENTER
smart key.
Heat unit to 350
°
C.
Press
EDIT
smart key.
Press
ENTER
smart key.
SET POINT:
will appear on display.
Enter 350.
Press
CLR
keypad button.
Use the number pad on the right side of the keypad to enter
350.
Press
ENTER
.
Press
ESC
key on keypad twice.
Press
SAVE
smart key. The system will now ramp to 350
°
C.