LCR7000 & RHEOSIGHTS P/N: 974194 |
Rev: A
|
ECO: 52468
www.dynisco.com
From lab to production,
providing a window into the process
-80-
Allow unit to stabilize at 350
°
C for 45 minutes.
Calibrate unit at 350
°
C.
Press
SYS
Key and
CAL
smart key.
Select RTD Sensor in the Calibrate Select mode.
Press
once.
Press
SELECT
smart key.
Select
DOUBLE
smart key.
Press
STABLE
smart key.
Press
CLR
key on keypad.
Enter reading displayed on the BTP 1000A display.
Press
CLR
.
Use the number pad on the right side of the keypad to enter the temperature. Enter the
numbers in order as they appear on the BTP 1000A display, also pressing the decimal
where necessary.
Note
: Numbers are entered by just pressing the appropriate number key.
Press
ENTER
smart key.
Record OFFSET and GAIN numbers. Press SAVE smart key.
Temperature calibration verification.
Allow machine to re-stabilize at 350
°
C. for 15 minutes. Record temperature.
Change temperature set point to 300
°
C. Repeat steps 2.13.1 through 2.13.5 replacing
350
with
300.
Allow machine to re-stabilize at 300
°
C for 45 minutes. Record temperature.
Change temperature set point to 230
°
C. Repeat steps 2.13.1 through 2.13.5 replacing
350
with
230
.