LCR7000 & RHEOSIGHTS P/N: 974194 |
Rev: A
|
ECO: 52468
www.dynisco.com
From lab to production,
providing a window into the process
-44-
RheoSights offers corrected data based on Rabinowicth correction and Bagley correction.
Rabinowitch correction can be performed on a single test or multiple tests (different
approaches). Bagley correction has to be performed on more than one tests which meet
certain criteria (see Appendix for details). Furthermore, Bagley/Rabino corrected Shear
Viscosity and Extensional Viscosity/Stress/Rate can be plotted based on the corrections.
If a single test is selected, Rabino Corrected Shear Viscosity can be plotted. Simply select it
on for left axis or right axis.
If two or more tests that meet the criteria of Bagley correction are selected, the indicator on
the right of the “Special Graphs” select box will turn to green, which means all the
correction-based data are unlocked and can be plotted.
6.
Corrections