Most of the rheometer comes pre-assembled to your door, however certain parts are prone
to breakage if they were placed in their normal operating position during shipping. These
items will need to be installed before safe operation of the rheometer is possible. These
items include the Load Cell (models 7000 & 7001), Interface cables for the Load Cell and
Pressure Transducers (models 7001 & 7002), Cable connection for the Edge Device, printer
(if applicable), Laser Micrometer (if applicable), and AC Power connections.
Keep the wooden crate in storage in case you ever need to move or ship the rheometer
back to Dynisco or to another location.
First find the surge protector/power box. This is an extension cord with many sockets
available for use. Plug it in and make sure the power is OFF (the ON switch should not be
lit). There are a number of power cords supplied with the rheometer. The thickest
(heaviest) of these is used for the rheometer. Connect this thickest cord to the back of the
rheometer base and into the isolated filter bank #3 in the power box. Power to the PC and
monitor should connect to the isolated filter bank #1. The printer (if used) can be
connected to any remaining outlet.
The data processing system consists of an Edge Device and the RheoSights software
package. Connecting the interface cable, from the back of the rheometer, to any USB port on
the Edge Device, makes the link between the Edge Device and the rheometer. On the back
panel of the rheometer, you will find a female DB9 connector for a serial port connection.
If you have purchased the HAMA Laboratories PA02L High Resolution Laser Micrometer
(available for models 7000, 7001, and 7002), you will find a digital I/O board, and interface
cable, and the laser micrometer itself. The digital I/O board base address switch is preset
at 2E0h. If this address is already assigned to some other device in your computer, you
must set this switch for a different address. Reference the HAMA Laboratories User’s
Guide.
3.
Unpacking the Rheometer
4.
Power Connection
5.
Edge Device Connection
6.
Laser Micrometer Connection