Data Print
If it is desired to print out photometric values at a fixed wavelength interval
during spectral scanning, set the Text Print parameter to Peak, Valley or All
Data. Otherwise, set this parameter to Not Print.
List Interval (nm)
If the Text Print parameter is set to ON, specify a wavelength interval at
which photometric values are printed out.
This parameter is specifiable within a range of 0.1 to 100.0 nm in
increments of 0.1 nm.
4. System Setup
Press <4> to select system setup from the Time Scan Sub Menu and a sub menu
list will appears on the right side of the screen as shown in the illustration below.
These parameters are the same as 3.3 except Peak Threshold.
Refer to the System Setup of Wavelength Scan in Section 3.4.
1
Auto
2
D2 Only
3
WI Only
11/10/02
14:25
600.0
nm
0.854
ABS
Time Scan
1
Parameter Setup
1
Auto
2
Sample Setup
2
Lamp Change WL (nm)
350.0
3
Print Setup
3
WI Lamp
On
4
System Setup
4
D2 Lamp
On
5
Save Parameter
5
Peak Thershold
0.001
0
Forward
6
Baseline
System
7
Response
Medium
8
Init Delay (s)
0.1
Select Function: _
Lamp Change Mode
3.5.2 Performing a Time Scan measurement
On completion of measurement conditioning setting, press the <0> key and the
measurement screen appears on the LCD. When the conditioning screen is
switched over to the measurement screen, the instrument checks the user-
specified parameters.