11/04/02
9:54
600.0
nm
0.854
ABS
Wavelength Scan
1
Parameter Setup
1
Data Mode
ABS
2
Sample Setup
2
Start WL (nm)
600.0
3
Print Setup
3
Stop WL (nm)
500.0
4
System Setup
4
Up Scale
3.000
5
Save Parameter
5
Low Scale
-3.000
6
User Baseline
6
Scan Speed (nm/min)
400
0
Forward
7
Num Cycles
1
8
Cycle Time (s)
0
9
Display Format
Sequential
Select Function: _
1
%T
2
ABS
3
E(S)
4
E( R)
Data Mode
Select one of the following modes:
a) %T (percent transmittance measurement)
b) ABS (absorbance measurement)
c) E (S) (Energy on the sample side measurement)
d) E (R) (Energy on the reference side)
Start WL (nm)
Start Wavelength for wavelength scanning operation. Enter a long wavelength
value of the measurement wavelength range.
This parameter is specifiable within a range of 190 to 1100 nm in increments of
0.1 nm.
Stop WL (nm)
Stop Wavelength for wavelength scanning operation. Enter a short wavelength
value of the measurement wavelength range.
This parameter is specifiable within a range of 190 to 1100 nm in increments of
0.1 nm.
Note: 1. The minimum scan width is 10 nm.
2. The data read interval varies depending on the wavelength range.
3. The data reading wavelength interval varies depending on the
specified wavelength range.
3-21