%T: Within a range of 0 to 600.0 in increments of 0. 1.
Response
Three response speeds are available: Fast, Medium and Slow.
Most application requires a “Medium”. For better reproducibility, choose
the slow response parameter. For high-resolution measurement, select
“Fast” response parameter.
Note:
The data reading time interval varies depending on the specified scan time.
Determine a period of scan time according to the purpose of measurement. As
the scan time increase, data is smoothed more to reduce noise.
Scan Time (sec)
Reading Interval (sec)
Scan time < 100
0.1
100
≤
Scan time < 200
0.2
200
≤
Scan time < 500
0.5
500
≤
Scan time < 1000
1.0
1000
≤
Scan time < 2000
2.0
1000
≤
Scan time < 5000
5.0
5000
≤
Scan time < 9999
10.0
Scan time
≥
10000
100.0
2. Sample Setup
Press <2> to select sample setup from the Time Scan Sub Menu and a sub menu
list will appears on the right side of the screen as shown in the illustration below.
These parameters are the same as 3.3.
11/14/02
11:26
600.0
nm
0.854
ABS
Time Scan
1
Parameter Setup
1
Name
SAMP-1
2
Sample Setup
2
Path Length (nm)
10.0
3
Print Setup
4
System Setup
5
Save Parameter
0
Forward
Input Name:_
3-36