11/10/02
9:47
600.0
nm
0.854
ABS
Time Scan
1
Parameter Setup
1
Data Mode
ABS
2
Sample Setup
2
WL (nm)
600.0
3
Print Setup
3
Scan Time (s)
500
4
System Setup
4
Up Scale
3.000
6
Save Parameter
5
Low Scale
-3.000
0
Forward
Select Function: _
1
%T
2
ABS
3
E(S)
4
E(R)
Data Mode
Select one of the following modes:
a)
%T (percent transmittance measurement)
b)
ABS (absorbance measurement)
c)
E (S) (Energy on the sample side measurement)
d)
E (R) (Energy on the reference side)
WL (nm) (Wavelength)
Enter a wavelength to be used for measurement. DB-20 has a range of 190
to 1100 nm in increments of 0.1 nm.
Scan Time
Specify a scanning time which corresponds to the upper limit value on the
abscissa axis of spectral graph.
Range of 60 to 99999 seconds in increments of one second.
Up Scale (Upper Scale)
Upper limit on Y axis.
Range for Y-axis upper limit in each Data Mode is as follows:
ABS: Within a range of –3.000 to 3.000 in increments of 0.001.
%T: Within a range of 0 to 600.0 in increments of 0. 1.
Low Scale (Lower Scale)
Lower limit on Y-axis.
Range for Y-axis upper limit in each Data Mode is as follows:
ABS: Within a range of –3.000 to 3.000 in increments of 0.1.